For dealers, trainers and studio teams
A DICOM setup is finished when the family can use the files
Use this free handoff checklist to test the complete path from an ultrasound machine to a browser gallery. A successful Verify test is only the first step.
Published by SonoLink · October 24, 2026 · No signup required
Download the printable checklist (PDF)Use it for a clear handoff
The transfer checks are useful when a dealer installs a machine, a trainer teaches a studio team or an owner changes image-sharing service. The worksheet includes SonoLink-specific acceptance steps and can be adapted to another receiving service. It is not a diagnostic validation, security certification or legal-record retention plan.
Use a test record without real patient information. Confirm the clinic’s approved transport and data-handling arrangement before any patient data is sent. Do not put passwords, patient identifiers or private gallery links in the worksheet or a public support post.
1. Record the configuration
| Record | Why it matters |
|---|---|
| Exact machine and software version | Menus, cine controls and licensed options differ by release |
| DICOM Storage/PACS option | A brand or model listing does not prove the installed option is enabled |
| Remote AE Title, host and port | These belong to the receiving studio’s account, not the machine’s local identity |
| Connection and transfer trigger | Record Ethernet/hotspot and manual/on-save/end-exam behavior |
| Original destination and fallback | Keep the previous configuration available during the pilot |
For SonoLink, copy destination values from that studio’s dashboard. Do not copy another clinic’s AE Title. Confirm model/software guidance on the machine page; the Samsung HM70 EVO guide distinguishes field experience from manufacturer documentation.
2. Prepare representative test media
- Save a new still after applying the intended settings.
- Save a short cine with visible motion; 3–5 seconds is useful for an initial test, within the unit’s limits.
- Test a rendered 4D cine separately if the studio uses it.
- Check image compression and cine inclusion separately where the unit offers those controls.
- Do not treat an ordinary JPEG export or proprietary raw volume as proof of a moving DICOM clip.
Keep resolution and frame rate unchanged for the first test. An old saved clip may retain its original encoding. Confirm unfamiliar controls with the manufacturer or service provider.
3. Verify every delivery stage
- Connection: Verify/Test succeeds for the correct receiving service.
- Transfer: send the test still and cine, then check the machine queue. Connection success alone does not establish transfer.
- Receiving study: confirm the correct record, visible still and completed moving-video playback.
- Family access: create a selected-media link and apply the intended PIN and expiration settings.
- Separate phone: open the link outside the staff account, play the full clip and download both items.
- Downloaded files: open the photo and play the saved video with motion. Record pass/fail separately for a 4D clip.
- Staff repeat: ask the receiving operator to perform the workflow using the intended transfer trigger and actual connection.
Record send-start, queue-finish and playable-video times separately. A fast short test is not a promise for every exam. Note whether a failure occurred during transfer, conversion or viewing so support can investigate the right stage.
4. Finish with ownership and a fallback
Record who accepted the setup and when, the steps that passed, any open issue and the studio’s support contact. Explain how to select the previous destination if the pilot fails. Keep the earlier service until the new workflow and the old-record access plan are both established.
Historical galleries do not move automatically. Use the switching and export guide before cancelling a prior subscription. SonoLink’s current scope is elective, non-diagnostic sharing with US hosting; confirm any other practice or country requirements separately.